Support for the Kirkby Microwave 85033 SMA calibration and verification kit on Anritsu VNAs

General information about the 85033 SMA calibration and verification kit

For general information about the kit, not specific to Anritsu, see the 85033 support page.

Supported Anritsu instruments

The 85033 calibration and verification kit can be used on all Anritsu VNAs, although not the Sitemaster, as the Sitemaster has no way to enter any coefficients of a calibration kit.

File downloads for Annritsu VNAs

We don't have files for download on our website at the minute, but we can easily create them for the VectorStar, Lightning and ShockLine VNAs. Just ask, and tell us what model you have. If you don't have a VectorStar, Lightning and ShockLine VNAs, then the coefficients will need to be entered manually. The coefficients are given below.

Coefficients for the 85033 kit in the format used by Anritsu

The following should be sufficient to set up any modern Anritsu VNA. Should you require any further information, then contact us.

Standard label: 85033. Serial number xxx
Type C0 C1 C2 C3 Offset length (mm)
Female short         12.41
Female open 53.004 135.167 95.6305 30.937 11.93
Male short         17.34
Male open 50.389 761.36 -968.22 68.376 16.52

Over the operating frequency range of this kit, the inductance of the short is negligable, so one should set L0=L1=L2=0